dialog W 2 CL2/PR R (20 mg)
dialog W 2 CL2/PR R
dialog W 2 CL2/PR R
Double-channel measuring and control instrument for free chlorine and pH-Value - additionaly ORP optional
dialog W 2 CL2/PR R (20 mg)
- Double measurement of free chlorine and pH -optionally with additional ORP measurement
- Easy and safe operation by plain text menu guidance
- Log book function
- Graphical trend display of the last 3 h
description
The dialog W 2 CL2 PR R is a sophisticated instrument for the precise measurement and control of the chlorine concentration and the pH-value - additionally an ORP measurement is optional. It is available in a wall mounted housing.applications
Drinking Water / Beverages
Waste Water Treatment
Disinfection
particular characteristics
- Variable assignment of relay and current outputs
- 2 limit values with delay
- Dosage check
- Semi-automatic two point-calibration for pH
- Slope and zero point monitoring for pH
- Automatic or manual pH-value compensation for pH
- Serial interface RS 485
- 2 or 3 scaleable, galvanically isolated 0/4 .. 20 mA outputs
- 2 separate PID-controller
- Real time clock
- Background illuminated two-line LC-Display
- Password function
information
Measuring and control instruments of the dialog series are also available for 117 VAC and 24 VDC (only wall mounted housing) - wihtout additional charge.technical data
measuring parameter |
|
| free Chlorine | 0 .. 4.00 mg/l |
| 0 .. 20.00 mg/l | |
| pH-value | 0 .. 14.00 pH |
| ORP | -1500 .. +1500 mV |
input characteristics
| accuracy | 1 % mg/L Cl2 of range, +/- 0,01 pH, +/- 1 mV | ||||
| temperature measuring range | -30 .. +140°C | ||||
| temperature coefficient | non-linear | ||||
| conditions |
|
output characteristics
| output signal | 3 0/4 .. 20 mA (scaleable, galvanically isolated) | ||||
| load | max. 500 Ohm | ||||
| registration range | Free scaleable within the measuring range | ||||
| voltage output | +/- 6 VDC for impedance converter | ||||
| serial interface | RS 485 | ||||
|
power supply
| line voltage | 24 / 117 / 230 VAC, +6 / -10 %, 40 .. 60 Hz optional: 24 VDC |
| power consumption | 10 VA |
ambient conditions
| ambient temperature | operation 0 .. +50°C, storage -20 .. +65°C | ||
| relative humidity | max. 90% rH at 40°C (non-condensing) | ||
| protection class |
|
controller
| control response | On/off controller (adjustable hysteresis) P/PI/PID-controller (pulse-pause, pulse-frequency or continous controller output) |
| relays | 3 relays, each with potential-free CO contact, max. 250 V, 6 A, 550 VA |
| onset delay | 0 .. 200 sec till controller active |
| digital input | Controller stop via external contact |
certificates and approvals
| CE-symbol | Declaration of conformity: The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives. The manufacturer confirms successful testing of the product by affixing the CE symbol. |
mechanical construction - wall mounted housing
| material | ABS |
| dimensions | 165x160x85 mm |
| weight | 1.25 kg |
| connection | Spring-loaded terminals |
dialog W 2 CL2/PR R
order information
name |
description |
article number |
| dialog W 2 CL2/ PR R | wall mounted housing, 230 VAC, 0 .. 4 mg/l | 115212K |
| dialog W 2 CL2/ PR R 20 mg | wall mounted housing, 230 VAC, 0 .. 20 mg/l | 115230K |
accessories
accessories
| CHECKIT Comparator | The CHECKIT Comparator is a compact handy colorimeter unit which is suitable for both mobile and static analysis work. CHECKIT Comparator is a colorimetric rapid test for easy determination of Free Chlorine, Total Chlorine, Chlorine Dioxide or Ozone concentrations that can be used as an indication measurement for our amperometric measuring systems. Naturally, regarding resolution and accuracy, the rapid test does not come close to the Krypton systems, and it cannot replace a fotometric measurement. However, it provides a quick orientation and can therefore prove helpful during installation or routine comparison. The CHECKIT Comparator comes in a robust plastic casing with all necessary equipment. |
| PKV 30-DPS | Converter for PROFIBUS-DP for max. 32 instruments of the K 100 oder dialog series. Converting of RS 485 to PROFIBUS-DP interface, instrument in a housing for installation on a top hat trail, dimension 105 X 105 x 80 mm (HxBxT), connectioncable inclusive |
| S-341 data logger | The bus data logger S-341 on the base of the RS-485 with Kuntze protocol reads, shows, and stores measuring data of up to 10 Kuntze instruments with up to 8 parameters each. |
![]() |
Download datasheetAll information about this product for your reference |
|
Collect and zip downloads
If you want to download several documents, you can collect and download them as one zipped file to save time |
Download collector
There are no files in your download-collector