Measurement and control instrument for dissolved oxygen
K 100 W K100
- Easy and safe operation by plain text menu guidance
- Safety by password function
description
The K 100 (W) O2 is a sophisticated single-channel instrument for measuring and control dissolved oxygen. Additionaly temperature measurement of the measuring media is possible. The instrument is available in a panel mounted housing (K100 O2) and in a wall mounted housing (K 100 W O2).applications
Waste Water Treatment
Drinking Water / Beverages
Food Industry
particular characteristics
- 2 limit values with delay, assigned to alarm relay
- 2 separate PI-controllers
- Dosage check
- Background illuminated two-line LC-Display for measured value and temperature
- Serial interface RS 485 optional
- Scaleable, glavanically isolated 0/4 .. 20 mA output
- Display of relay status and error messages
- Invertible alarm output, permanent or pulse contact
technical data
measuring parameter |
|
| dissolved Oxygen | 0.0 .. 20.0 mg/l |
input characteristics
| temperature measuring range | -30 .. +140 °C |
output characteristics
| output signal | 0/4 .. 20 mA (scaleable, galvanically isolated) | ||||
| load | max. 500 Ohm | ||||
| registration range | free scaleable within the measuring range | ||||
| serial interface | RS 485 (optional) | ||||
|
power supply
| line voltage | 24 / 117/ 230 VAC, +6/-10%, 40 .. 60 Hz |
| power consumption | 10 VA |
ambient conditions
| ambient temperature | Operation 0 .. +50 °C, storage -20 .. +65°C | ||||
| relative humidity | max. 90% rH at 40°C (non-condensing) | ||||
| protection class |
|
controller
| control response | on/off controller (adjustable hysteresis) P/PI controller (pulse-pause, pulse-frequency or continuous output) |
| relays | 2 relays, each with a potential-free NO contact, max. 250V, 6 A, 550 VA 1 alarmrelay with potential-free CO/NO contact, max. 250V, 6A, 550 VA |
| onset delay | 0 .. 200 sec till controller active |
| digital input | Controller stop via external contact |
certificates and approvals
| CE-symbol | Declaration of conformity: The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives. The manufacturer confirms successful testing of the product by affixing the CE symbol. |
mechanical construction - panel mounted housing
| material | Noryl |
| dimensions | 90x90x116 mm |
| installation dimensions | 92x92x140 mm |
| weight | 0.75 kg |
| connection | push-srew terminals |
mechanical construction - wall mounted housing
| material | ABS |
| dimensions | 165x160x85 mm |
| weight | 0.95 kg |
| connection | Spring-loaded terminals |
K 100
K 100 W
order information
name |
description |
article number |
| K 100 O2 | Dissolved oxygen, panel mounted housing, 230 VAC | 100350K |
| K 100 W O2 | Dissolved oxygen, wall mounted housing, 230 VAC | 105350K |
accessories
accessories
| PKV 30-DPS | Converter for PROFIBUS-DP for max. 32 instruments of the K 100 oder dialog series. Converting of RS 485 to PROFIBUS-DP interface, instrument in a housing for installation on a top hat trail, dimension 105 X 105 x 80 mm (HxBxT), connectioncable inclusive |
| S-341 data logger | The bus data logger S-341 on the base of the RS-485 with Kuntze protocol reads, shows, and stores measuring data of up to 10 Kuntze instruments with up to 8 parameters each. |
![]() |
Download datasheetAll information about this product for your reference |
|